Rh sets logtype2 (/lg1 option), Xe applies basic settings, Ye applies basic settings (cold reset) – Yokogawa Value Series FX1000 User Manual

Page 128

Advertising
background image

3-46

IM 04L21B01-17EN

QA

Sets the number of mantissa

digits to display (/LG1 option)

Syntax

QA p1<terminator>
p1 Number of digits to display (2, 3)

Query

QA?

Example Set the number of mantissa digits to display to 2.

QA2

RH

Sets LogType2 (/LG1 option)

Syntax

RH p1
p1 LogType2 (Linear, Pseudo)
Linear: Sets LogType2 to log linear input
Pseudo: Sets LogType2 to pseudo log input

Query

RH?

Example Set LogType2 to pseudo log.

RHPSEUDO

WF

Sets the Modbus connection

limitation (/C7 option)

Syntax

WF p1<terminator>
p1 Modbus connection limitation (USE, NOT)

Query

WF?

Example Place limitations on Modbus connections.

WFUSE

WG

Sets an IP address that is

allowed to connect through the

Modbus interface (/C7 option)

Syntax

WG p1,p2,p3<terminator>
p1 Registration number (1 to 10)
p2 Registration (ON, OFF)
p3 IP address (0.0.0.0 to 255.255.255.255)

Query

WG[p1]?

Example Allow connections from IP address

192.168.111.24. Use registration number 1.
WG1,ON,192.168.111.24

Description You can use this command when Modbus

connection limitations have been placed (through
the use of the WF command).

WP

Sets the phase, wiring system,

and input voltage (/PWR1 option)

Syntax

WP p1,p2<terminator>
p1 Phase and wiring system

Single-phase two-wire:

1P2W

Single-phase three-wire:

1P3W

Three-phase three-wire:

3P3W

p2 Input voltage (120V, 240V)

Query

WP?

Example Set the FX to use the single-phase two-wire

system with an input voltage of 120 V.
WP1P2W,120V

Description • When parameter p1 is set to 3P3W, parameter

p2—the input voltage—is fixed to 240V.

• If you change these settings, the VT ratio, CT ratio,

and integration low-cut settings will all be initialized.

XE

Applies basic settings

Syntax

XE p1<terminator>
p1 Storage of settings (STORE, ABORT)

Example Save the basic settings.

XESTORE

Description To apply settings that you have changed using

the basic setting commands, you need to save
the settings using the XE command. Be sure to
save the settings using the XE command before
you change the execution mode from the basic
setting mode to the operation mode. Otherwise,
the new settings will not be applied.

YE

Applies basic settings (cold

reset)

Syntax

YE p1<terminator>
p1 Application of settings

STORE Saves the basic settings and

restarts the instrument

ABORT Restarts the instrument without

saving the basic settings

Example Save the basic settings and restart the

instrument.
YESTORE