Index – Yokogawa PC-Based MX100 User Manual

Page 105

Advertising
background image

4-17

IM MX100-01E

Specification

3

2

1

4

5

Index

The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).

Input

Measurement

Range Type

Rated Measurement

Range

Measurement Accuracy

Highest

Resolution

(1 Digit)

Integration Time: 16.67 ms

or More

Integration Time: 1.67 ms

DC voltage

60 mV (high res.)

0.000 to 60.000 mV

±(0.05% of rdg + 20 digits)

±(0.1% of rdg + 100 digits)

1 µV

1 V

–1.0000 to 1.0000 V

±(0.05% of rdg + 2 digits)

±(0.1% of rdg + 10 digits)

100 µV

6 V (high res.)

0.0000 to 6.0000 V

±(0.05% of rdg + 20 digits)

±(0.1% of rdg + 100 digits)

100 µV

Thermocouple

(excludes RJC accuracy

, when

burnout is OFF)

PLATINEL

0.0 to 1400.0°C

±(0.05% of rdg + 1°C)

±(0.1% of rdg + 4°C)

0.1°C

PR40-20

0.0 to 1900.0°C

±(0.05% of rdg + 2.5°C)

Except 300 to 700°C: ±6°C

less than 300°C: not guaranteed

±(0.1% of rdg + 12°C)

Except 300 to 700°C: ±25°C

less than 300°C: not guaranteed

NiNiMo

0.0 to 1310.0°C

±(0.05% of rdg + 0.7°C)

±(0.1% of rdg + 2.7°C)

WRe3-25

0.0 to 2400.0°C

±(0.05% of rdg + 2°C)

Except 0 to 200°C: ±2.5°C

2000°C or higher:

±(0.05% of rdg + 4°C)

±(0.1% of rdg + 7°C)

Except 0 to 200°C: ±12°C

2000°C or higher:

±(0.1% of rdg + 11°C)

W/WRe26

0.0 to 2400.0°C

±(0.05% of rdg + 2°C)

Except 100 to 300°C: ±4°C

less than 100°C: not guaranteed

±(0.1% of rdg + 8.5°C)

Except 100 to 300°C: ±12°C

less than 100°C: not guaranteed

N (AWG14)

0.0 to 1300.0°C

±(0.05% of rdg + 0.7°C)

±(0.1% of rdg + 3.5°C)

XK GOST

–200.0 to 600.0°C

±(0.05% of rdg + 0.5°C)

Except –200 to 0°C:

±(0.2% of rdg + 0.7°C)

±(0.1% of rdg + 2.5°C)

Except –200 to 0°C:

±(1% of rdg + 2.5°C)

Measurement interval, integration time, and filter:

Measurement Interval Integration Time

Filter

Rejected Noise and Notes

500 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

1 s

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply

frequency and set 16.67 or 20 ms

2 s

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

5,10, 20, 30, 60 s

100 ms

Rectangular

10 Hz and its integer multiples

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate especially

with temperature measurement. In such cases, set the measurement interval to 1 s or higher, or use

the 4-CH, High-Speed Universal Input Module or the 10-CH, Medium-Speed Universal Input Module.

Reference junction compensation:

Switch between external and internal on each channel, remote
RJC function available

Reference junction compensation accuracy:

When measuring temperature greater than or equal to 0°C and
when the temperature of the input terminal is balanced

Type R, S, W: ±1°C

Type K, J, E, T, N, L, U, XK GOST: ±0.5°C

Type N(AWG14), PLATINEL, NiNiMo, WRe3-25, W/WRe26: ±1°C
Note: The internal reference junction compensation is fixed to 0°C

for type B and PR40-20

Maximum input voltage:

DC voltage at 1-V range or less, thermocouple, and DI (contact):

±10 VDC (continuous)

Other measurement ranges: ±120 VDC (continuous)

Normal-mode voltage:

DC voltage, thermocouple, DI (LEVEL): 1.2 times the range rating
or less (50/60 Hz, peak value including the signal component)

Normal-mode rejection ratio:

For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ±0.1%)

For integration time of 1.67 ms: 50/60 Hz is not rejected

4.6 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30) Specifications

Advertising