IS 320 • IGA 320 Manual

General 13

becomes diffuse. The targeting light only marks the center of the measuring spot, not the size.
A measuring tape can also be used to determine the distance between the object and
pyrometer.
If the pyrometer is supplied with power, the LED targeting light is automatically active. If the
targeting light does not activate, check whether the pyrometer is properly connected to the
power supply (consult the window pyrometer parameters in Chapter 4 for more information).

It is possible to connect a push button to the pyrometer to switch the targeting light on and off
(see Section 3.1.1, Pin assignment of the male socket on the back of the pyrometer).

3.3 Optics

The pyrometers are equipped with one of the following optics. Optics focus to a certain
distance, i.e. at these distances each optics achieves its smallest spot size in relation to the
measuring distance. The spot size will change at any other distance (shorter or longer).

Note: The measuring object must be as least as big as the spot size.

The drawing and the table below illustrate the size of the spots in mm in relation to the
measuring distance. Values between the data shown below can be calculated by interpolation.

The spot size for measuring distance 0 is the aperture diameter “D” of the optics.

Type Optics

Temperature range

a : M

*

)

a

M

a

1

M

1

a

2

M

2

D

IS

32

0

4

550 ... 1400 °C

(MB 14)

83 : 1

250

3

500

15

1000

39

9

600 ... 1600 °C

(MB 16)

125 : 1

2

500

13

1000

35

650 ... 1800 °C

(MB 18)

192 : 1

1.3

500

12

1000

33

5

550 ... 1400 °C

(MB 14)

92 : 1

600

6.5

1000

17

2000

43

600 ... 1600 °C

(MB 16)

133 : 1

4.5

1000

14

2000

36

650 ... 1800 °C

(MB 18)

188 : 1

3.2

1000

11.3 2000

32

6

550 ... 1400 °C

(MB 14)

93 : 1

1400

15

2000

26

3000

43

600 ... 1600 °C

(MB 16)

156 : 1

9

2000

17

3000

30

650 ... 1800 °C

(MB 18)

200 : 1

7

2000

14

3000

26

IG

A

32

0

4

300 ... 1300 °C

(MB 13)

125 : 1

250

2

500

13

1000

35

9

400 ... 1800 °C

(MB 18)

208 : 1

1.2

500

11.4 1000

32

5

300 ... 1300 °C

(MB 13)

133 : 1

600

4.5

1000

13.5 2000

36

350 ... 1650 °C

(MB 16.5)

188 : 1

3.2

1000

11.4 2000

32

400 ... 1800 °C

(MB 18)

231 : 1

2.6

1000

10.3 2000

30

6

300 ... 1300 °C

(MB 13)

156 : 1

1400

9

2000

16.8 3000

30

400 ... 1800 °C

(MB 18)

233 : 1

6

2000

12.4 3000

24

Spot size [mm]

Measuring distance [mm]

D

a

a

1

a

2

M

2

M

1

M