8 troubleshooting – LumaSense Technologies IS 12-Al/S User Manual

Page 31

Advertising
background image

IS 12 AI & IS 12-AI/S

Troubleshooting 31

8 Troubleshooting

Before sending the pyrometer for repair, try to find the error and to solve the problem with the

help of the following list.

Temperature indication too low

Incorrect alignment of the pyrometer to the object

New correct alignment to achieve the max. temperature signal

Incorrect focusing of the optics

Focus correctly to the measuring distance, if necessary change lens

Measuring object smaller than spot size

Change lens and measuring distance

Measuring object is not always in the measuring spot of the pyrometer (e.g. swinging

wire or pouring stream)

Use max. value storage or pyrometer with scanning system with max. value storage

Emissivity set too high

Set lower correct emissivity corresponding to the material

Lens contaminated or scratched

Clean lens carefully or replace it

Temperature indication too high

Emissivity set too low

Set lower correct emissivity corresponding to the material

The measurement is influenced by reflections of hot machine parts

Use mechanical construction to avoid the influence of the interfering radiation

(sighting tube)

Measuring errors

Indicated temperature is decreasing during the use of the pyrometer, contamination of
the lens

Clean lens. Recommendation: use of air purge

Indicated temperature is decreasing during the use of the pyrometer, although the air

purge unit is used. Probably compressed air is not clean or air failed

Clean the lens and use clean, dry and oil free compressed air

Air contamination in the sighting path between pyrometer and object

Change position of the pyrometer with a clean sighting path or use a ratio pyrometer

HF-interferences

Correct the connection of the cable shield

Instrument overheated

Use cooling jacket with air or water cooling

Temperature Indication is fluctuating, probably caused by changing emissivity

Wrong pyrometer type, use of ratio pyrometer recommended

Laser targeting light

Laser targeting light fails

Instruments temperature above 55 °C. Use cooling jacket.

Advertising
This manual is related to the following products: