1 general information, 1 information about the user manual, 1 legend – LumaSense Technologies M390S User Manual

Page 5: 2 safety, 1 additional m390s safety notifications

Advertising
background image

M390S Manual

General Information 5

1 General Information

1.1 Information about the user manual

Congratulations on choosing the high quality and highly efficient Mikron blackbody.
This manual provides important information about the instrument and can be used as a work of

reference for installing, operating, and maintaining your blackbody. It is important that you
carefully read the information contained in this manual and follow all safety procedures before
you install or operate the instrument.
To avoid handling errors, keep this manual in a location where it will be readily accessible.

1.1.1 Legend

Note: The note symbol indicates tips and useful information in this manual. All notes
should be read to effectively operate the instrument.

Attention: This sign indicates special information which is necessary for correct
operation.

Warnings and Cautions: The general warnings and cautions symbol signifies the
potential for bodily harm or damage to equipment.

1.2 Safety

This manual provides important information on safely installing and operating the
M390S blackbody. Several sections of this manual provide safety warnings to avert danger.
These safety warnings are specified with a warning symbol. You must read and understand the
contents of this manual before operating the instrument even if you have used similar
instruments or have already been trained by the manufacturer.
It is also important to continually pay attention to all labels and markings on the instrument and

to keep the labels and markings in a permanent readable condition.

Warning: The blackbody is only to be used as described in this manual. It is
recommended that you only use accessories provided by the manufacturer.

1.2.1 Additional M390S Safety Notifications

1. This M390S may not be used for any purpose other than the function for which it was

designed, namely, a high temperature blackbody radiance temperature source.

2. If the power cable requires replacement, replace only with equal, approved cable.
3. Cabinet covers may be removed while in operation if required for trouble shooting only.

DANGER -- High voltages and currents are present inside the enclosure.

4. Do not look into the blackbody cavity when temperatures are above 1100 °C. Temporary

flash blindness can occur at temperatures > 1200 °C assuming the eye blink response of
0.15s. Permanent eye damage could occur if you continuously stare at blackbody

temperatures about 1200 °C without protection. (Calculations suggest that permanent
damage would not occur with 0.15s blink exposure at 3000 °C.)

5. Do not attempt to operate with any of the safety features bypassed or disconnected.

Advertising