1 rs485 option – LumaSense Technologies M390 User Manual

Page 16

Advertising
background image

M390 Manual

Operation 16

4.5.1 RS485 Option

RS485 two wire is available as an option. The connection scheme is shown below:

RS485 for multi-drop applications

Each M390 address must be set to a different value.
To change the address of an M390 (if RS485/RS422 was ordered):

1. Press and hold PAGE button. ACCESS and Level 1 appear in lower display.

2. Press UP button until LEVEL 3 appears in lower display.

3. Press UP button until 3 appears and wait.

4. Press PAGE until COMMS H appears.

5. Press UP until COMMS J appears. (J COMMS is the RS-485 output)

6. Press SCROLL until ADDRESS appears.

7. Press UP or DOWN to set new address and wait.

8. Press PAGE until ACCESS appears.

9. Press DOWN until LEVEL 1 appears and wait.

Note: A CD containing a sample program is included with your order. For more details
on communications, refer to the electronic communication manual also found on the

CD provided.

M390 Unit 1
(Address 1)

M390 Unit 2
(Address 2)