Appendix B: Calibration Methods 30

Example 1: The HT (450 C) blackbody was factory calibrated to provide effective emissivity =

1.0 at 8-14 µm. You wish to use this source for a narrow band pyrometer at 7.5µm. At this
wavelength from the graph above, the blackbody exit radiance will be approximately equivalent
to a blackbody that has emissivity = 1.011. This will cause your pyrometer to receive 1.1% more

infrared energy causing it to read a higher temperature. The table below shows the expected
error. At 250 C, this pyrometer will read 251.4 C or 1.4 C high, so you should set the
blackbody setpoint to 248.6 C (250-1.4).

Normalized emissivity after 8-14um radiance calibration on the HT (450C) version

0.92

0.93

0.94

0.95

0.96

0.97

0.98

0.99

1

1.01

1.02