Chapter 1
_________________________________________________________________________________________

_____________________________________________________________________________

BE6021-11

1316A-2 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 9 of 97

Fig.1.1 Flow diagram of the 1316A-2 Multi Gas Monitor

1.2.2 NDIR Module and UA1372 Module (Optional) Measuring System

This block is the heart of the monitor. It contains the Infrared source

the sample cell and the Detector with specialized optical filters and

processing electronics necessary to do highly accurate measurement

of gas concentrations.

The UA1372 are connected to the NDIR module for data transfer to

the Controlling PC Software.

Measurement samples are sucked into the measurement chambers

via the Gas Inlet or the Zero Gas Inlet, depending upon the

command from the BZ6013. The internal gas-circuits are arranged

such that samples pass through the measurement chamber in a

continuous flow and directed from the NDIR sample cell to the

UA1372 Module to the Electrochemical Sensors. Exhaust gases are

expelled via the gas outlet. The Zero Gas used for zero calibration of

the NDIR Module and for Span Calibrating the Oxygen Sensor is

drawn via the Zero Gas inlet.