Function Check • 17

English

7. FUNCTION CHECK

 Perform functional check prior to each operation.

A complete functional check includes:

• Visual inspection for mechanical damage

• Visual and functional inspection of all movable pa-

parts

• Functional check of the control elements

CAUTION!

Risk of injury

If you notice any faults or deviations during the
functional check, do not use the unit.
 Have repairs of the unit carried out by seca

service or an authorized service partner.

8. TROUBLESHOOTING

9. SERVICING

The seca measuring rod does not require maintenance.

Should any part of it fail to function, please contact an

authorised service partner.

You can find your local service partner at

www.seca.com or just send an e-mail to

Problem

Cause/Solution

Measurement does not
return plausible results.

Measurement result read off incorrectly

- Use the correct edge for the measuring range

currently in use (“Performing length measure-
ment” from page 14)

- Read off the decimal places for the current

measuring range in the correct direction (“Per-
forming length measurement” from page 14)

Incorrect position/posture of patient

- Correct the position/posture of the patient

(“Performing length measurement” from
page 14)

Measuring rod mounted incorrectly on column
scale

- Mount the measuring rod correctly (“Before

you really get started…” from page 8)