Technical data, 1 general technical data – Seca 635 User Manual

Page 98

Advertising
background image

98 •

12.TECHNICAL DATA

12.1 General technical data

General technical data

seca 635

Dimensions of scale

• Depth
• Width
• Height

600 mm
640 mm

46 mm

Dimensions of weighing platform

• Depth
• Width
• Height

600 mm
600 mm

46 mm

Net weight

approx. 14.1 kg

Temperature range

• Operation
• Storage
• Transport

+10 °C to +40 °C / +50 °F to 104 °F

-10 °C to +65 °C / +14 °F to 149 °F
-10 °C to +65 °C / +14 °F to 149 °F

Air pressure

• Operation
• Storage
• Transport

700-1060 hPa
700-1060 hPa
700-1060 hPa

Humidity

• Operation
• Storage
• Transport

30 % - 80 % no condensation

0 % - 95 % no condensation
0 % - 95 % no condensation

Height of digits

25 mm

Power supply

Batteries, optional power pack

Power consumption

• with wireless module deactivated
• with wireless module activated

approx. 25 mA
approx. 42 mA

Maximum runtime in battery mode

• with wireless module disabled
• with wireless module activated

approx. 4,800 weighing operations

Power pack recommended

Medical device in accordance with
Directive 93/42/EEC

Class I with measuring function

EN 60601-1:

• insulated device, protection class II

• electrical medical device, type B

Advertising