Servicing – Seca 763 User Manual

Page 26

Advertising
background image

26

… the head stop slides down and does not stay in the required position?
– Place the scale carefully on its side so

that the base of the scale is accessible.

Insert the Allen key supplied in the rec-

tangular hole on the end of the measu-

ring rod as shown in the illustration and

tighten up the screw for the measuring

rod guide clockwise.

The measuring rod with the head stop

should remain freely mobile, but not be

able to slide down on its own.

Then stand the scale upright again.

8. Servicing

Your seca scale leaves the factory with an accuracy greater than ± 0,15 %. To ensure that

this accuracy is maintained, the product must be carefully set up and regularly maintained.

Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please

ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for

further information.

Advertising