Continuous display of measured result (hold), Perform relative measurements (zero) – Seca 264 User Manual

Page 60

Advertising
background image

60 •

7. Align the patient’s Frankfurt horizontal with the

matching one of the three Lines of the Frankfurt
measure.

8. Keep the brake button pressed and correct the

position of the head slide.

9. Read off the height on the head slide display.

10. Press the Enter key (send/print) to transmit the

height to receivers from the

seca 360° wireless

system.

– Press key briefly: send measured results to all

receive-ready devices

– Hold down: print measured results

Continuous display of

measured result (Hold)

If you activate the HOLD function, the measured value
continues to be displayed after the measurement is
complete. This allows you to move the head slide
before noting down the measured value.

1. Briefly press the arrow key (hold/zero), after you

have positioned the head slide.

The message “HOLD” is displayed.
The head slide can now be moved without the
measured value displayed changing.

2. To disable the HOLD function, press the arrow key

(hold/zero).
The “HOLD” message is no longer displayed.

Perform relative

measurements (zero)

The head slide display can be set to zero at any
position. This function allows relative measurements,
for example for measuring extremities.

1. Press the Start key

of the head slide.

“----” appears in the display.

2. Keep the brake button pressed and move the head

slide until a length value appears in the display.

3. Position the head slide at the point you want to set

the head slide display to zero.

4. Keep the arrow key (hold/zero) pressed until

the "ZERO" message is displayed.
The display is set to zero at the current position of
the head slide.