Before you get started … • 85

English

CAUTION!

Risk of injuries and damage to the device

The device must be tilted. The large overall
height of the device can result in injuries and
damage to the device.
− Ensure that there are no other persons in the

immediate vicinity.

− Ensure that there are no objects in the

immediate vicinity.

1. Tilt the device backwards.

ATTENTION!

Damage to the device due to excessive voltage

Standard power supplier units can supply a
higher voltage than is given on them. The scales
can overheat, catch fire,
melt or be short-circuited
− Only use genuine seca plug-in power supply

units with 9V or a controlled 12 V output
voltage.

2. Insert the connector plug of the power supply unit

in the connecting socket of the scales.

3. Carefully upright the device.

4. Plug the power supply unit into a mains socket.

5.6

Calibrate head slide

Before you can measure height using the device for the
first time, you have to calibrate it. You do so by
“teaching” the device a reference measurement.

Automated calibration

You can perform automated calibration using the
calibration rod supplied.

1. Press the Start key

of the head slide.

“----” appears in the display.