82 •

Installing the push-in

scale

ATTENTION!

Incorrect measurements due to installation
errors

If the push-in scale is installed incorrectly the
head slide cannot determine any usable
readings.
− Align the push-in scale so that the printing

remains visible when the scale is pushed into
the groove.

− Align the push-in scale so that the area which

shows the product number forms the bottom
end.

− Ensure that the push-in scale latches into

position under the end cap.

Push the push-on scale into the groove in the side
of the column elements until the scale latches into
position under the end cap.

5.3

Set up device

ATTENTION!

Incorrect measurement with force shunt

Correct measurements cannot be taken if the
patient's weight is not fully positioned on the
weighing area.
− Set up the device so that it only makes

contact with the ground via the foot screws.

− Set up the device so that the weighing area

does not touch any other objects.

− While you are weighing, ensure that the

patient does not touch the upright of the
device.

1. Place the device on a flat, stable surface.