AM1210 Type S Reference Standard Thermocouple
User’s Guide

2/4

1 Introduction and Specifications

1.1 Introduction

AM1210 Type S Reference Standard Thermocouple is constructed
using reference grade Platinum and Platinum vs. 10% Rhodium
alloys. It covers a temperature range of 0 °C to 1300 °C with short
term stability of 0.2 °C all the way to Freezing Point of Copper
(1084.62 °C). It is commonly used as reference standard to
calibrate industrial thermocouples. All thermocouple wires and
parts are specially cleaned and annealed before assembly. Every
AM1210 thermocouple is fully annealed and tested again to meet
the tolerance criteria as specified in the table below.

1.2 Specifications

Temperature Range

0°C to 1300°C

Type

Type S: Platinum/10 % Rhodium vs. platinum

Long Term Drift

±0.6°C at 1084.62 °C after 1 year typical usage

Tolerance (mV)

E(t

Cu

)=10.575±0.015

E(t

Al

)=5.860+0.37(E(t

Cu

)-10.575)±0.005

E(t

Zn

)=3.447+0.18(E(t

Cu

)-10.575) ±0.005

Short Term Stability

±0.2°C at 1084.62 °C

Diameter of thermocouple
wire

0.5 mm

Sheath Material

Quartz or Alumina

Sheath Dimensions

OD: 6 mm; Length: 500 mm

External Lead Wire

S type thermocouple wire, 500 mm

Protective Carrying Case

Included

2 Care and Handling Guidelines

1. DO NOT subject the thermocouple to any physical shocks

and vibrations.

a. When not using the thermocouple, keep it in a place

that’s not prone to drop, slam, bang, vibration or other

Advertising