Pressure / Process Calibration Equipment

Additel Datasheet 2014 / 2015

Corporate Headquarters

22865 Savi Ranch Parkway Ste F

Yorba Linda, CA 92887, USA

Salt Lake City Office

802 Bamberger Dr, Suite C

American Fork, UT 84003, USA

Phone: 714-998-6899

Fax: 714-998-6999

Email: [email protected]

Phone: 801-756-5570

Fax: 714-998-6999

Email: [email protected]

Environmental
Compensated Temperature: 14°F to 122°F(-10°C to

50

°

C )

Operating Temperature: 14°F to 122°F(-10C to 50C )

Storage Temperature: -4°F to 158°F (-20°C to 70°C )

Humidity: <95%

Pressure Connections
1/4NPT male, 1/4BSP male, M20X1.5 male

Other Connections available per request

Enclosure
Case material: SS

Wetted parts: 316L SS

Dimension: Ø1.3" x 4.84" (Ø33mm x 123mm)

Weight: 0.99 lb (0.4kg)

Warranty: 1 Year

Absolute Pressure

P/N

Pressure Range

Media

Accuracy(%FS)

Burst

Pressure

(psi)

(bar)

AP5

5

0.35

G

0.1

AP10

10

0.7

G

0.1

AP15

15

1.0

G

0.1

AP30

30

2.0

G

0.1

AP50

50

3.5

G

0.1

AP100

100

7.0

G , L

0.1

AP300

300

20

G , L

0.1

AP500

500

35

G , L

0.1

AP1K

1,000

70

G , L

0.1

AP3K

3,000

200

G , L

0.1

AP5K

5,000

350

G , L

0.1

SPECIFICATIONS

ORDERING INFORMATION

Model Number

Model

Range type:
PSI - range by psi
BAR - range by bar

Pressure port type:
N – 1/4NPT male
N2 - 1/2NPT male
B -1/4BSP male
M- M20 X 1.5 male

ADT160

GP300

PSI

N

02

Pressure range P/N:
See pressure range table

Accuracy:
02-0.025% of full scale
05-0.05% of full scale
10-0.1% of full scale

Accessories included
NIST Traceable Calibration Certificate

Optional Accessories

Model

number

Description

Picture

9060

Pressure module connection cable

Differential Pressure

P/N

Pressure Range

Media Accuracy

(%FS)

Burst

Pressure

Static

Pressure

Range

(inH

2

0) (mbar)

DP1

±1

±2.5

G

0.05

100×

±10 psi

DP2

±2

±5.0

G

0.05

100×

±10 psi

DP5

±5

±10

G

0.05

50×

±10 psi

DP10

±10

±25

G

0.05

20×

±10 psi

DP20

±20

±50

G

0.05

20×

±10 psi

DP30

±30

±75

G

0.05

20×

±10 psi

DP50

±50

±160

G

0.05

±10 psi

DP150

±150

±350

G

0.025

50 psi

DP300

±300

±700

G

0.025

50 psi

Additel 223A with ADT160 Pressure Module

* One year accuracy (including yearly stability) except DP1 and DP2

range which is 0.05%FS calibration accuracy and 0.05%FS yearly

stability.

Advertising