Automatic scanning, Action scope of buttons – Additel LogII User Manual User Manual

Page 6

Advertising
background image

6 / 12

1. Automatic Scanning

Once it starts, the software will automatically scan the connected UUTS without manual

operation. And the relevant UUT information frame will be displayed in the UUTs’ list area

(#2).

The real-time pressure value and curve of selected UUT will be displayed separately in

area #3 and #4 too.

If the software cannot detect the connected UUT, please check the following items.

Please check whether the serial drive or USB to RS232 drive are installed correctly.

Please make sure the cable is connected tightly with interface.

If users use USB to RS232 cable to connect with PC, please re-plug the USB port into
PC to make sure the cable could work properly.

Please check whether the UUT is turned on.

Please check whether the Baud rate of UUT matches the software.

Please make sure there is no other software of PC occupied the serial port used by
UUT.

2. Action Scope of Buttons

The functions of “Log”, “Setting”, “Clear”, “Unit” and “Export” are same with the five

buttons on the each frame of UUTs. And they are one to one correlate. As shown in picture

2-2.

The difference is the buttons on the toolbar could control all the connected UUTs,

however the buttons on each frame could only control the relevant one.

Advertising