Handheld Pneumatic Pressure Test Pump

User s Manual

'

Warnings and cautions

Low pressure gauges can very easily be over pressured if not careful.

Please take caution when applying pressures.

Do not exceed the safety pressure limit of 1000 psi (70bar).

This pump is not designed to be used in explosive or corrosive

environments.

Do not switch between pressure and vacuum mode under pressure.

Over tightening connectors may cause damage.

Store the pump in dry and non-corrosive environments.

Additel is not liable for any safety problems or damages caused by

misuse or incorrect operation.

All moisture and contaminants should be cleaned out of the liquid

isolator before creating a vacuum. This is accomplished by venting

the pump when at high pressure.

For the least risk in contamination of the reference gauge, the DUT

should be installed to the pressure port which is closest to the vent

valve.

Pressure range:

Temperature:

Humidity:

Adjusting fineness:

Safety pressure:

Pressure media:

Size:

Weight:

95 vacuum to 375 psi 25 bar

0~50

< 85%RH

10 Pa ( 0.1 mbar )

< 1000

70

Air

Height: 5.12'' ( 130 mm );

Base: 9.45'' ( 240 mm ) x 4.72'' ( 120 mm )

3.3 lb ( 1.5 kg )

%

(

)

psi (

bar )

Specification

1

2 3

4

5

6

7

Configuration & Air routine

1

Quick connector

-

3

7

4

5

2

Air routine

Troubleshooting

2 - Vent valve

clockwise to close/

to open)

counterclockwise

3 - Isolation valve

isolates the calibration volumn from pressurizing system)

4 - Hand pump

5 - Fine adjust andle

h

(clockwise to increase pressure)

6 - Security bolt pin

(use it to ensure safety)

7 - Pressure/vacuum (P/V) selector

Postive pressure: pull it out; Vacuum: push it in)

Problem

Cause

Solution

Hand pump will not generate pressure

It is difficult to increase pressure

Hard to use the fine adjust

It is difficult to turn the quick connectors

The isolation valve is not open

B. The O ring seal is loose or damaged

A. The isolation valve is not closed

C. The O-ring seal is damaged.

E. The connector is mismatched to the gauge pressure port

A Too much force was previously applied

.

.

.

-

.

.

.

A. Vent valve is not closed.

C. P/V selector is in the wrong position

B. The gauges are not tightened.

D. The thread surface is not smooth

F. Dust or contaminants are blocking the pressure flow

B. The threads have no lubrication

.

.

.

Open the isolation valve

Replace the O ring seal

Close the isolation valve

Use the correct adapter

Secure the quick connection with less force

.

.

Replace the O ring seal

-

.

.

.

-

.

Close the vent valve.

Pressure pull the P V selector out; Vacuum push the P V selector in

Tighten the reference gauge or the gauge under test

Use Teflon tape on the thread and turn it in tight.

Repeat pressurizing the pump several times and then suddenly vent which will push any contaminants out

Apply lubrication to the threads

:

/

:

/

.

.

.

.

ADT914

[Version number:1405V11]