Additel 920 Pneumatic High Pressure Test Pump User Manual User Manual

Adt920, Pneumatic high pressure test pump user manual 's, Warnings and cautions air routine

Advertising
background image

7

3

2

5

8

Warnings and cautions

Air routine

Pneumatic High Pressure Test Pump User Manual

's

[Version number:1407V02]

Troubleshooting

Configuration & Size

ADT920

1

2

5

8

7

9

3 4

6

1

Quick connector

-

8 - Vent valve

(

)

clockwise to close/

to release pressure

counterclockwise

7 - Isolation valve

solates the calibration volume from the hand pump and

check valve

(i

)

2 - Hand pump

5 - Fine adjust handle

(Turn Clockwise to increase pressure)

4 - Pressure/Vacuum lock

3 - Pressure/vacuum (P/V) selector

(Positive pressure: pull it out; Vacuum: push it in)

Do NOT switch while unit is under any pressure

9 - Vent port

6 - Maintenance cleaning port

17.12 ''

2

'

1.26'

54

10.63''

5.50''

19.92''

Low pressure gauges can very easily be over pressured if not careful.
Please take caution when applying pressures.
Do NOT exceed the safety pressure limit ( 3500 psi / 235 Bar ).
This pump is not designed to be used in explosive or corrosive environments.
Do not switch between pressure and vacuum mode under pressure. Make sure you
have vented the pressure and opened the isolation valve before switching to vacuum.
Over tightening connectors may cause damage.
Pump should be stored in a dry/clean environment.
Additel is not liable for any safety problems or damages caused by misuse or
incorrect operation.
Do NOT hold pressure above 1000 psi for more than 4 hours.
All moisture and contaminants should be cleaned out of the liquid isolator before
creating a vacuum. This is accomplished by venting the pump when at high pressure.
For the least risk in contamination of the reference gauge, the DUT should be
installed to the pressure port which is closest to the vent valve.

Specification

Pressure range:

Temperature:
Humidity:
Adjustment resolution:
Overpressure Safety Limit:
Pressure media:
Weight:

95% vacuum to 3000 psi ( 200 bar )

( Based on an atmospheric pressure of 1 bar )

℃ /

(

psi ( 235 bar )

0 to 50

32 to 122 F

< 85%RH

0.001psi 0.1mbar

10 Pa )

3500

Air

14.4 lbs ( 6.5 kg )

RemarkIf local atmosphere pressure is 1 bar, the vacuum can reach to -0.95 bar;

If local atmosphere pressure is P, the vacuum can reach to - 0.95P bar.

Problem

Cause

Solution

Hand pump will not generate pressure

It is difficult to increase pressure

Hard to use fine adjust

Not easy to turn fittings

The isolation valve is not open

B. The O-ring seal is loose or damaged

D. Internal tubing is dirty.

B. The gauges are not tightened.

D. The connector type is mismatched to the gauge pressure port.

A. Too much force was previously applied.

.

C. The connector type is mismatched to the gauge pressure port.

A. Vent valve is not closed.

C. P/V selector is in the wrong position

A. The isolation valve is not closed.

C. The O-ring seal is aged or frayed.

E. Air route is blocked/dirty

B. The threads have no lubrication

.

.

.

Open the isolation valve

Replace O-ring seal.

Disassemble and clean (Request diagram).

Tighten the reference gauge and / or the gauge under test

Use the correct adaptor.

Do not over tighten. Hand tight is sufficient.

.

.

Use the correct adaptor.

Close the vent valve.

Pressure : Pull P/V selector out; Vacuum : Push the P/V select or in.

Close the isolation valve.

Pressurize to high pressure and release pressure quickly via the vent valve

Lubricate the thread

Replace O-ring seal.

.

.