Remark, Preparation basic operation – Additel 927 Hydraulic Pressure Test Pump User Manual User Manual

Page 2

Advertising
background image

Remark:

A Additel has made a concerted effort to provide complete and current information for the proper use of the equipment. The product specifications and other

information contained this manual are subject to change without notice.

B

:

: Above pictures are just for reference.

A

Open the reservoir

The media level should be 2/3 the height
of the reservoir.

B

Filling/emptying the reservoir

Clean any impurities from the media and reservoir.

C

Lock the reservoir

Preparation

Basic Operation

Connection

A

Vent

B

Zeroing should
be operated
under this
condition only.

①open

②open

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

Pressure Gauge

Pre-pressure

E

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

Pressure Gauge

①open

②open

④turn it to the

mid position

③turn it to

greatest

degree

①open

②open the upper cover

Fill or empty media

①close the upper cover

②close

Vacuum preparation

G

0

0.02

0.04

0.06

0 .08

- 0.1

Vacuum Gauge

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

②open

①open

④turn it to the

mid position

③turn it to the

smallest

degree

Vacuum process

H

0

0.02

0.04

0.06

0 .08

- 0.1

Vacuum Gauge

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

③fine adjust

①close

②turn it to approach

the desire value

SW V mA COM 24V

SW V mA COM 24V

SW V mA COM 24V

dial gauge

pressure switch

pressure transmitter

k

Pressure Gauge

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

flexible hose

Pressurizing process

F

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

Pressure Gauge

①close

①close

If necessary,
please generate
high pressure
using #4 handle

③generate high-pressure

and fine adjust pressure

②Pre-pressurization

Purge process

C

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

①close

③turn it to the smallest degree

②Turn all the

way out

Purge process (cont.)

D

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

②open

①open

If air is still trapped in the pump
then repeat the purge process 2
or 3 more times until all the gas
is removed from the system.

,

③Turn all the way in

bubble

Additel Corporation.USA

www.additel.com

If it is difficult to generate
vacuum pressure then
repeat the purging
process to make sure all
the gas is removed from
the system.

,

Vacuum displayed on the gauge

Wait for gauge to vent to zero