Model 2200R

Page 19

5.15 Status

5.15.1 The 2200R continuously checks the integrity of all stored data and monitors the

condition of the measuring system. If a fault is detected, the STATUS LED above
the display will change from green to red. When this occurs, enter the menu and
scroll through the menu until the STATUS LED illuminates. The display will show a
numerical code to assist you, in correcting the problem.

The following table shows the display codes, causes and remedies:

CODE POSSIBLE CAUSE

SUGGESTED REMEDY

0

Normal Condition

No action required

1

ORP probe efficiency <60% or
>110% Nernst coefficient (Probe
slope out of tolerance

Clean probe and recalibrate. Salt
bridge / probe may need
replacement

2

Temperature Sensor off Scale

Check for open or short
connections of temperature sensor
input (Terminals BK & YL on TB2)

3

ORP reading off Scale

Verify process. Check for open or
short connections

4*

Memory Loss

Call your AquaMetrix Rep. Or
Water Analytics directly

5*

EPROM write failure

Call your AquaMetrix Rep. Or
Water Analytics directly

6

Calibration out of limit

Check scale setting (See range
change in Utility Menu)

7

Factory settings in force as a result
of the “Factory Reset Procedure”

8

ORP calibration not correct, or not
performed

Re-calibrate (See Sec. 5.2 & 5.3)

9

Temperature calibration not correct
or not performed

Calibrate temperature (See Sec.
7.4)



NOTE: Code 4 and 5 could be serious failures so the alarm relay will activate, in addition to the

status LED changing to red, if programmed by the position of DIP switch No. 5 of Bank S1.
Refer to Section 4.7.