Model as conductivity cells – Aquametrix AS Series Conductivity Sensors User Manual

Page 2

Advertising
background image

Model AS Conductivity Cells

www.WaterAnalytics.net

978-749-9949

Toll free - 855-747-7623 (855-pHProbe)

Cell Constants

0.01, 0.1, 0.5, 1.0, 10 and 50

Thread size

3/4” NPT

Temperature Compensation*

Automatic, 0 to 105

0

C

Cable Length

20 ft. (6.1 m)

Insertion Length

Variable. See table below

Wetted Materials

316 stainless steel, polypropylene,

Epoxy and EPDM. Compression

fitting: CPVC or 316 stainless steel.

Safety Feature

Shoulder on cells provides

an integral safety stop

Temperature and Pressure

3 metres (10 ft.) coaxial, terminated

with a spade lug for active electrode,

other wires tinned
316 stainless steel, polypropylene,

epoxy and EP

Temperature and Pressure

Type

Max pres. (psi)

Max temp. (

0

C)

P

50

60

100

50

150

20

S

200

105

Cell Constant Selection

Constant

Code

0.01

001

0.1

01

0.5

05

1.0

1

10

10

Temperature Compensa-

tion Device Selection

Description

Code

PT1000 RTD

1

3 kΩ ohm NTC

2

Type Selection

Description

Code

Cell with PVC fitting

P

Cell with 316 stainless steel fitting

S

Cell for Ball Valve Assembly

HT

MS Series of probes
MHMSC, 24 inch (610 mm) long 1” CPVC pipe w/ cable clamp
MS-HTC, 3/4” socket-weld CPVC ball valve assembly
AM3422, AM3444 and AM3455 316 SS body conductivity sensors

Technical Data

Dimensions

Ordering Information

Related Products