Technical data, Order information, Related products – Aquametrix AM3455 Series Conductivity Sensors User Manual
Page 2
Model AM3455 Sanitary Conductivity Sensors
Dimensons
www.WaterAnalytics.net
978-749-9949
Toll free - 855-747-7623 (855-pHProbe)
Technical Data
Wetted Materials
Insulator
FDA approved DuPont PTFE TeflonTM®
O-rings
FDA approved Parker EJ150-75 (3077) EPDM
Electrodes
Electrodes - 316L Stainless Steel
Maximum Pressure / Temperature
100 PSIG at 120°C
Cell Constants
0.01 / 0.05 / 0.1 / 1.0 / 10.0
Connections
Process Fitting
#16AMP solid end cap by Tri-Clover
Electrical
Stripped and tinned leads
Temperature Compensation
3KΩ NTC Thermistor or Pt1000
Flange Size
A (±0.1)
1.5”
5.25”
2”
5.25”
2” w cell constant 10
9.2”
Lengths are tailored for use with a Tri-Clover #7MP TEE, position sensor horizontally.
Order Information
MS Series of epoxy body conductivity probes
Related Products
Cell Constant Selection
Constant
Code
Constant
Code
0.01
01
1.0
1
0.05
005
10
10
0.1
01