Hygrocontrol manual for type 86 ex page – COSA Xentaur Hygrocontrol Type 86 Ex User Manual

Page 10

Advertising
background image

Transmitter: Transmitters built according to EN 50020 as intrinsically safe
supply instruments. Intrinsically safeness is given by Zener Barriers which
are encapsulated together with the connector for the cable to the probe
and therefore may not be manipulated by a user. These Barriers separate
the intrinsically safe power supply for the probes from the internal and
external current and voltage loops which are not intrinsically safe.

Concept of Realisation

HYGROCONTROL Manual for Type 86 Ex Page

9

Probe Type 861

Conception for Humidity Detectors Type 86-EX

Transmitter Type 86

Connection Cable

(intrinsically safe)

Power Cable

Signal Cable

EX - Environment (Zone 1)

Category 2 GD

-20°C <= ta <= +60°C

EX - Environment (Zone 0)
Category 1 GD

-40°C <= ta <= +120°C

30 mbar < P < 30 bar

No EX -

-20°C <= ta <= +40°C

Environment

but intrinsically safe
Supply for Probes 861

Advertising