COSA Xentaur COSA 707 User Manual

Page 10

Advertising
background image

COSA Instruments

USER MANUAL COSA 707

www.cosaxentaur.com

10

Draft and temperature are measured at the tip of the sampling probe.

5.1 Gas flow diagram

Gas flow diagram with CO purging pump

Position Description

1

Sampling probe

2

Triple hose

3

Condensate separator

4

Star filter

5

Non return valve

6

Gas pump

7

Sensor chamber

8

O

2

-Sensor

9

2. sensor

10

3. sensor

11

4. sensor

12

5. sensor

13

CO Purge pump

14

Pressure sensor *

* optional

5.2 Electrochemical measured principle

The oxygen content of the sample gas is measured with a 2 electrode electrochemical
sensor.

Toxic gases like carbon monoxide (CO), nitrogen oxide (NO), nitrogen dioxide (NO

2

), sul-

phur dioxide (SO

2

), and hydrogen sulphide (H

2

S) are measured with 3 electrode sensors.

The electrochemical sensors are based on gas diffusion technology.

Advertising