Appendix b: flow diagram of set-up state user inte, Xdt user’s manual appendices 37 – COSA Xentaur Xentaur XDT User Manual

Page 44

Advertising
background image

XDT User’s Manual Appendices

37

Appendix B: Flow Diagram of Set-Up State User Interface

Hi/Lo

SET

SET

SET

Lo

Lo

Lo

Hi

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

Test Analog Output by forcing it to
high, low or midpoint.

(E. G. if Span is 4 to 20, Hi =20mA, Lo=4mA, Mid=12mA)

Set measured attenuation at a known low dewpoint

Set low dewpoint for above attenuation

Calibration Midrange Adjust.

Set dewpoint corresponding to low end of analog output

Factory default is low range of sensor, e. g. -100˚C for XTR-100

Factory default is high range of sensor, e. g. +20˚C for XTR-100

When locked, instrument parameters can not be changed.

Factory default is locked.

Set dewpoint corresponding to high end of analog output

Lock or unlock the instrument.

Cable Length Compensation.

blank ˚F ˚C ppm LBS G/m3 PSI % ˚C˚F Temp˚F Temp˚C

Choose alternate units to show when displaying dewpoint.

Choose sensor type

XTR-100

XTR-65

measure

attenuation of

sensor

if no buttons pressed for
3 seconds shows the
dewpoint for a moment.

SET

Select Span of Analog output: 0 to 24mA or 4 to 20mA

Select behavior of Analog output during self calibrations or sensor wakeup

NOR

-

means analog output tracks measurement

FR

-

means analog output freezes

Factory default is blank, (there is no alternate display).

CAUTION! These are set properly at the factory

with the shipped sensor. Do not modify these

settings, unless you are changing the sensor.

˚C

˚C˚F

DEWPOINT

value

˚C

value

˚C

CNF/CBL/OPN

CNF/CBL/OPN

CNF/CBL/SHR

OPN

OPN

SHR

END

UN/LOC

A.0./RNG/

A.0./RNG/

A.0./TST

Hi/Lo

A.0./OPN

A.0./SPN/

0.24 OR 4.20

CAL/ADJ

-1.15X

-80.0

DB/

-16.24

SEN/XT1

ALT/DSP

XT1 XT6 DB

LOC

appears when changing
settings of locked unit

Resets the

instrument

Open cable at BNC connector

Open cable at BNC connector

Short cable at BNC connector

Displayed only if Analog Output option is installed

Power ON initialization
and sign on message

4 min anywhere without
key press

{

sensor type

{

changes by 0.01 db

{

changes by 0.01 units

{

changes by 0.1˚C

{

changes by 0.1˚C

changes by 0.1˚C

{

units

{

selection

SET

A.0./
NOR OR FR

{

selection

Power-Up with

MODE

key depressed

{

Un-

Lock

Lock

Hi

Lo Mid

Test

Current Loop

Output

OK

Circuit

Open

Legend: A slash '

/

' in the display area, is used to depict two alternately shown (flashing back and forth) messages.

denotes beeping. Buttons without designator or arrow leading out, perform no function

go to
Operating
State

Advertising