GW Instek GDS-1000A-U Series firmware upgrade guide User Manual

How to upgrade the firmware

Advertising
background image

How

to Upgra

de the Firmwa

re

1

How to Upgrade the

Firmware

This

applic

ation note describes how

to

manually

up

grade the GD

S-

1000

-U/GDS-1

000

A-

U’s f

irmwa

re.

Required equipments

USB fla

sh drive

A stan

dar

d

U

SB

flash

drive

with the belo

w

firmw

are

file

s copied

into

the root direc

tory

Firmware

Three file

s

ar

e required:

dso1kh.rom

dso1ks.rom

dso1kmon.ro

m

Front panel overview with items related to the u

pgrade

GDS-1

000-

U/GDS-1

000A

-U

A

pplica

tion N

ote

2

Procedure

1. Inserting

the

USB fla

sh

drive into the

USB ho

st port

Insert the U

SB flash dr

ive

includ

ing n

ew

firmware

fi

les i

nto

the GDS-1

000

-U

/GDS

-1

000

A-U’s

USB

host port.

2. T

urning

t

he

oscillo

scope

on

Power the osci

lloscope

up

an

d

at

the same ti

me,

press the

F1 k

ey seve

ral times. The o

scilloscope

should

au

toma

ti

cal

ly boot i

n the fi

rmwa

re

upgrade mod

e as in

the sn

apshot below

.

3. Up

grading t

he

firmware

Press the F

4

ke

y. The

os

cilloscope auto

matically

st

art

s up

gr

ad

ing

th

e firm

w

are.

When the st

at

us

in

di

ca

tor

at the bottom

of

the

display

show

s the

“Upgr

ade S

u

cces

s!!”

message,

upgra

d

ing

th

e f

irmwar

e i

s

complet

ed.

(ongoing)

(comple

ted)

4. C

onfirming the

upgrade

Power the o

scilloscope o

ff

and on

to reb

oot the

system. Open

the sy

stem

in

format

ion

by

pressin

g

the Utility ke

y

S

ys

tem Inf

o (F

4)

.

Ut

ility

S

yst

em

Inf

o

.

Check the fir

m

ware ver

sio

n in the

syste

m

inform

at

ion

sc

reen t

hat

it

is

being

upd

at

ed.

Upda

tin

g

the f

irm

wa

re is co

m

pleted

Advertising