GW Instek GSP-810 Software Manual User Manual

Page 3

Advertising
background image

Copyright©2003 GOODWILL Instrument CO., LTD.

(c) When the converting ends, the following windows will show up. Click “OK” to quit.

(d) The test.txt file contains the date and time of when making the SPF file, center frequency,

RBW, reference level, span and 501 points of data. Since it is the record of measurement, therefore

more then one measured data could be stored. It dependents on how long the trace records. In this

case, there are four groups of data.

5. Point Data

The Point data is 10 bits DAC value. Equation for converting DAC value to dBm reading is as

follow.

Ref.Level

DAC

reading

dBm

+

×

=

80

1

023

,

1

For example, the point data is 942, reference level is –20 dBm, the reading of the point in dBm is

)

(

33

.

26

20

33

.

6

)

20

(

80

1

023

,

1

942

dBm

=

=

+

×

2003.05.12

Date

Time

Ref.
level

RBW

Center

Freq.

Span

501 points data

2nd

3rd

4th

Advertising