GW Instek GDS-800 Series User Manual User Manual

Page 50

Advertising
background image

GDS-800 Series Digital Storage Oscilloscope Operation Manual

50

z

No Go When

/

: Select the judgment condition of violation for

“Go-No Go” function.

z

No Go When

: If the subject signal is not violating the templates,

the system will judge such condition as “No Go” situation by selecting this

function.

z

No Go When

: If the subject signal is violating the templates, the

system will judge such condition as “No Go” situation by selecting this function.

z

More: Press F5 softkey to other utilities.

CAL. OUTPUT: The adjustable signal output provides a range for a signal of

probe compensation which can be adjusted from 1kHz to 100kHz, 1kHz per step;

and duty cycle is adjusted from 5% to 95%, 5% per step. This function can

satisfy user with the probe compensation more easily and accurately. NOTE:

This function is available for GDS-806S/C and GDS-810S/C only

Figure 5-24: The menu for adjustable probe compensation signal

z

Wave Type: Press F1 soft key to select adjustable probe compensation

signal or another two output signals for demonstration.

: A demonstration signal which designed for the purpose of long record

length.

: A demonstration signal which designed for the purpose of peak detects.