HT309

EN - 9

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL CHARACTERISTICS
Accuracy is referred to the following reference conditions: temperature 23°C ± 5°C with
relative humidity < 70%RH.

The instrument is calibrated for a standard white light source consisting of an incandescent
lamp, at a temperature/colour of 2856°K in compliance with Class A.

Calibration for LED sources (with spectrum response different from that of standard white
light) is based on the calibration for standard white light source in Class A using the
appropriate correction factors (see Table 1)

Illuminance measurement (Autorange)

Full scale (Lux)

40

400

4000

40k

400k

Resolution (Lux)

0.01

0.1

1

10

100

Accuracy ±

(3%reading)

Full scale (Fc)

40

400

4000

40k

Resolution (Fc)

0.01

0.1

1

10

Accuracy ±

(3%reading)

NOTE 1: 1Fc=10.76Lux, 1Klux=1000Lux, 1Kfc=1000Fc

NOTE 2: for temperature/colours different from the reference one, accuracy becomes 6%reading

Distance range for luminous intensity measurement: 0.01 ÷ 30.47m / 0.01 ÷ 99.99ft

Accuracy on angular deviation from cosine characteristic

30° ±

2%

60° ±

6%

80° ±

25%


6.1.1. Spectrum

response

The spectrum response of the photodiode with filter is almost identical to the CIE photo-
optical curve V(

λ) as shown in the following graph.

0

10

20

30

40

50

60

70

80

90

100

380

440

500

560

620

680

740

V(?)

HT307


Fig. 2: CIE curve V(

λ).

6.1.2. Sensor
The sensor is a silicon photodiode with filter on the spectrum response.

Relative sensitivit

y

(%

)

Wavelength (nm)

HT309

V(λ)