Technical specifications – HT instruments HT326 User Manual

Page 18

Advertising
background image

HT326

EN - 17

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL

FEATURES

Accuracy is indicated as [%reading + (number of digits*resolution)] at 23°C±5°C, < 75%HR

DC Voltage

Range Resolution Accuracy

Input impedance

Overload

protection

400mV 0.1mV

(0.8%rdg + 3dgt)

10M

600V DC/AC rms

4V 0.001V

(0.8%rdg + 2dgt)

40V 0.01V

400V 0.1V

600V 1V

(1.0%rdg + 2dgt)


AC Voltage

Range Resolution

Accuracy

(40

400Hz)

Input impedance

Overload

protection

400mV 0.1mV

not

declared

10M

600V DC/ACrms

4V 0.001V

(1.0%rdg + 3dgt)

40V 0.01V

400V 0.1V

600V 1V

(1.2%rdg + 3dgt)


DC Current

Range Resolution Accuracy

(*)

Voltage

drop

Overload

protection

10A 0.01A

(1.2%rdg + 3dgt)

200mV Fusibile

10A/600V

(*) Accuracy is guaranteed for current up to: 6A continuous measurement, 7A 3 minutes uninterrupted

measurement, up to 10A 2 minutes continuous measurement.


AC Current

Range Resolution

Accuracy (*)

(40

400Hz)

Voltage drop

Overload

protection

10A 0.01A

(2.0%rdg + 5dgt)

200mV Fusibile

10A/600V

(*) Accuracy is guaranteed for current up to: 6A continuous measurement, 7A 3 minutes uninterrupted

measurement, up to 10A 2 minutes continuous measurement.


Resistance

Range Resolution Accuracy

Max Open

Circuit Voltage

Overload

protection

400

 0.1

(1.0%rdg + 8dgt)

circa 0.4V

600V DC/AC rms

<30sec

4k

 0.001k

(1.0%rdg + 2dgt)

40k

 0.01k

400k

 0.1k

4M

 0.001M

40M

 0.01M

(2.0%rdg + 2dgt)



Advertising