HT8051

EN - 7

4.2.5.

key

In operating modes SOUR mA, SIMU mA, OUT V and OUT mV pressing this key allows
setting the output current/voltage with automatic ramp, with reference to measuring ranges
0

 20mA or 4  20mA for the current and 0  100mV or 0  10V for the voltage. Below

shows the available ramps.

Ramp type

Description

Action

Slow linear ramp

Passage from 0%

100% 0% in 40s

Quick linear ramp

Passage from 0%

100% 0% in 15s

Step ramp

Passage from 0%

100% 0% in steps of 25%

with ramps of 5s

Table 1: List of available ramps for output current/voltage

Press any key or turn off and then on again the instrument to exit the function.

4.2.6. Adjuster

knob

In operating modes SOUR mA, SIMU mA, OUT V and OUT mV the adjuster knob (see
Fig. 1 – Position 8) allows programming the output current/voltage generated with
resolution 1

A (0.001V/0.01mV) / 10A (0.01V/0.1mV) / 100A (0.1V/1mV). Proceed as

follows:

1. Select operating modes SOUR mA, SIMU mA, OUT V or OUT mV.
2. In case of current generation, select one of the measuring ranges 0

 20mA or 4 

20mA (see § 4.2.7).

3. Press the adjuster knob and set the desired resolution. The arrow symbol “

” moves to

the desired position of the digits on the main display following the decimal point.
Default resolution is 1

A (0.001V/0.01mV).

4. Turn the adjuster knob and set the desired value of output current/voltage. The

corresponding percentage value is indicated on the secondary display.

4.2.7. Setting measuring ranges for output current
In operating modes SOUR mA and SIMU mA it is possible to set the output range of
generated current. Proceed as follows:

1. Switch off the instrument by press the

key

2. With 0-100% key pressed switch on the instrument by press the

key

3. The value “0.000mA” or “4.000mA” is shown at display for approx. 3 seconds and then

the instrument back to the normal visualization


4.2.8. Adjusting and disabling the Auto Power OFF function
The instrument has an Auto Power OFF function which activates after a certain period of
idleness in order to preserve the instrument’s internal battery. The symbol “ ” appears on
the display with enabled function and the default value is 20 minutes. To set a different
time or deactivate this function, proceed as follows:

1. Press the “

” key to switch on the instrument and, at the same time, keep the MODE

key pressed. The message “PS – XX” appears on the display for 5s. “XX” stands for
the time indicated in minutes.

2. Turn the adjuster to set the time value in the range 5

30 minutes or select “OFF” to

disable the function.

3. Wait 5s until the instrument automatically quits the function.