1. With the instrument in stand-by, press the

key to

select the desired type of reference (see § 4.2)

2. Press the

key three times to enter the indirect 3-

point measuring section. The symbol “

“ appears

on the display with side “1” flashing

3. Position the instrument in the lowest point (1) of

measurement (see Fig. 11 ) and press the

key

or the side MEAS key. The result appears on the

first partial display. The symbol “

“ appears on the

display with side “2” flashing

4. Position the instrument as horizontally as possible

(2) (see Fig. 11) and press the

key or the side

MEAS key to measure. The result appears on the

second partial display. The symbol “

“ appears on

the display with side “3” flashing

5. Position the instrument in the highest point (3) of

measurement (see Fig. 11) and press the

key

or the side MEAS key to measure. The result
appears on the second partial display

6.

The final value of the result obtained with the
combination of previous measures (see dashed line
in Fig. 11) is shown on the resulting display

7.

Measuring result is automatically saved in the
instrument's memory

Fig. 11: Indirect 3-point measurement – Mode 2