10 TECHNICAL SPECIFICATIONS

This product conforms to the prescriptions of the European directive on low voltage
73/23/EEC (LVD) and to EMC directive 2004/108/CE.

10.1 CHARACTERISTICS
CURRENT MEASUREMENT

Range current

Resolution

Accuracy

2

÷ 2500A

1A

±(2% rdg +2 digit)

¾

Additional error due to not centering:

2% reading

¾

Crest Factor max:

3

¾

Frequency (Hz):

50

±6%, 60±6%

¾

Conversion mode:

TRMS

¾

Bandwidth: 3200Hz

¾

Sampling frequency:

64 samples each 20ms

¾

Memory size:

1Mbyte

¾

Sampling period:

1s, 6s, 30s, 1min, 5min

¾

Serial port:

RS232


10.1.1 Safety standards
• Comply with:

IEC / EN 61010

• Insulation:

Class 2, double reinforced insulation

• Pollution:

Level

2

• Max

height:

2000m

• Overvoltage category:

CAT III 1000V

∼ to ground

CAT IV 600V

∼ to ground


10.1.2 General data
Mechanical characteristics
• Size:

120 (L) x 80 (W) x43 (D) mm

• Max diameter cable clamped:

174mm

• Head

length:

600mm

• Weight (including battery):

about 0.5kg

• Protection index:

IP65 (RS232 connector Closed)


Power supply
• Battery type:

2 batteries 1.5V LR6-AA-AM3-MN1500

• Battery life:

> 6 months (with charged batteries)


10.2 ENVIRONMENTAL CONDITIONS
10.2.1 Climatic conditions
• Reference temperature:

23 ± 5°C

• Operating

temperature:

-20°

÷ 60°C

• Operating humidity:

up to 100% RH (RS232 connector closed)

• Storage temperature:

-20 ÷ 60 °C

• Storage humidity:

up to 100% RH (RS232 connector closed)


10.2.2 EMC
This instrument was designed in accordance with EMC standards in force and its
compatibility has been tested in accordance IEC / EN61326.

EN - 13