Field Deployment

Program the Instrument

In order to set up a log or download data, you must connect the instrument to a
computer running Win-Situ 5 Software, or a RuggedReader Handheld PC running Win-
Situ Mobile. See page 51.

Position the Instrument

Place the instrument at the desired depth. Position the instrument below the lowest
anticipated water level, but not so low that the pressure sensor range might be
exceeded at the highest anticipated water level. Refer to the tables below to determine
usable depth.

A BaroTROLL Instrument can be deployed with a non-vented
instrument to compensate level data for atmospheric pressure
data. Make sure the clocks on both instruments are synchronized,
and install the BaroTROLL in a location that will never be
submerged in water.

See page 40.

Range

Effective Range

Usable Depth

PSIA

PSIA

kPA

Meters

Feet

30

15.5

106.9

10.9

35.8

100

85.5

589.5

60.1

197.3

300

285.5

1968

200.7

658.7

500

485.5

3347

341.3

1120

1000

985.5

6795

692.9

2273

* At sea level (14.5 PSI atmospheric pressure)