For best results, perform a 2-point calibration at these points before going into the
field.

100% calibration point:

Fill bubbler cup with clean water. Turn on the pump, regulate with the pinch

a.

valve. Insert the TROLL 9500 into the bubbler, resting the RDO adapter
plate on the rim of the cup. Allow 5-10 minutes for temperature and oxygen
equilibration in the water. Select RDO in the software Navigation tree, click
the

Calibrate button, then click Run.

B1

When readings have stabilized, click

b.

Next. Continue with the Zero Oxygen

Point for a 2-point calibration.

0% calibration point:

Fill the orange-based cal cup with sodium sulfite

a.

(Na

2

SO

3

) solution and slide it up around the RDO

sensor until the cap and thermistor are completely
submerged. Be sure that there are no air bubbles
on the foil.
Select

b.

Run in the Zero Oxygen Point screen. When

readings have stabilized, click

Next, then Finish to

finalize the calibration.

Rinse

c.

very thoroughly after calibrating in sodium

sulfite.

The RDO sensor is now calibrated and ready to use.

Turbidity sensor:

The sensor has been factory-calibrated over its full range (0
to 2000 NTU). If field calibration is required, we recommend a zero point calibration.
Use deionized water and the clear flow cell. Make sure that the 3-way valve on the
inflow port is closed.

In the office: Assemble the site data

6

The software wizard will prompt for the following information for the output report:

• Well diameter & depth

• Pump model & type

• Tubing type

• Screen length • Depth to water, top of screen & pump intake
• Tubing inner diameter and length (for software volume calculations)

Install batteries

1

Install D-cells in the TROLL 9500

a.

: Remove the battery

compartment cover, insert batteries, replace cover, and tighten.
Install and charge the RuggedReader battery pack:

b.

This process is illustrated in the RuggedReader Quick Start.
To calibrate 100% dissolved oxygen, install alkaline D-cells in the

c.

portable air pump as shown on the diagram inside the lid.

Connect the cable

2

Remove the red dust caps. Insert the TROLL 9500 connector

a.

into the cable connector. Push, twist, and click to lock.

The

“click” ensures that the cable is securely attached.

Connect the TROLL Com to the other end of the cable.

b.

Install software on your desktop/laptop PC from the

3

In-Situ

®

software CD

Win-Situ

®

4

Microsoft

®

ActiveSync

®

: Click on Microsoft Tools, select ActiveSync 4.X.

Win-Situ Sync

TIP: Although Pocket-Situ 4 is pre-installed on the RuggedReader. You need
ActiveSync and Win-Situ Sync on the PC to transfer files between the computers.

In the office: Launch the software

4

Connect the TROLL 9500 to the PC and launch Win-Situ 4. Click

a.

Next through 4

screens of the Connection Wizard to set up the COM port.
Select the Com Port, tap

b.

Find.

TIP: Calibration and low-flow setup can optionally be done on the RuggedReader using
Pocket-Situ 4.

In the office: Calibrate the water quality sensors

5

pH, ORP, and conductivity sensors

Remove the hydration bottle from the pH/ORP sensor.

a.

Save it to

keep the sensor hydrated between uses.
Use the orange-base cal cup with cutout and indentation. Fill with

b.

Quick Cal, slide it up around the sensors, including the RDO cable.
In Win-Situ, tap

c.

Parameters, tap QuickCal, follow the Wizard.

TIP: Detailed steps for calibration and low-flow software are in the MP TROLL 9500

Operator’s Manual on the software CD.

RDO cable outside cup
RDO sensor inside cup

Low-Flow with

RDO

®

Optical Dissolved Oxygen

Quick Start

Bottom Layer of Case

Multi-Parameter
TROLL

®

9500

with RDO, Conductivity,
Turbidity (optional),
& pH/ORP sensors—
Remove hydration
bottle from pH/ORP
sensor before use!

Quick Cal
Solution

6’
Cable

Air pump

Flow
cell

Cal cup

for Quick Cal

Bubbler

contains
batteries

Hardware fittings

for flow cell

Alkaline batteries

for TROLL 9500

Top Layer of Case

RuggedCable

TROLL

®

Com

TROLL

®

9500

Flow cell

RDO cable inside cup,

RDO sensor outside cup

12 V adapter

for RuggedReader PC

AC adapter

for RuggedReader

®

PC

RuggedReader

®

battery

Win-Situ
software CD

and Quick Start Guides

TROLL

®

Com

serial comm cable

USB cable

for RuggedReader PC

RuggedReader

®

handheld PC

TROLL 9500

maintenance kit

Low-Flow with RDO

Quick Start Guide

Low Flow Base Plate

(not pictured)