In-Situ Level TROLL Data Loggers Care & Maintenance Guide User Manual

Instruction sheet, Care and maintenance of aqua troll, And level troll

Advertising
background image

Instruction Sheet

Care and Maintenance of Aqua TROLL

®

and Level TROLL

®

Instruments

User Maintenance

Overview

It is important for users to perform scheduled
maintenance on their instruments to sustain the
accuracy and longevity of the probes and cables. The
frequency of this maintenance depends on the
characteristics of the deployment site, including
humidity levels and the degree of fouling.

Users should be aware of the conditions at their
deployment sites and develop appropriate maintenance
schedules to replace desiccant, clean the instruments,
and send in the instruments for factory calibration.
Users should check instruments often during the first
portion of the deployment to determine the frequency of
maintenance. General maintenance should be
performed as often as possible. Factory service should
be performed at least every 12 months.

Preventing Condensation

Desiccant

Desiccant protects electronics from condensation,
which can cause irreparable damage and loss of data.
Indicating desiccant changes from blue to pink as it
becomes saturated with moisture.

Desiccant stages (from left)
New, nearly expired (replace now), expired

It is extremely important to use a properly-sized
desiccant for your deployment, and to change
desiccant often
. Desiccant should be changed before
the entire volume has turned pink, and you should use
enough desiccant to effectively keep cables,
instruments and electrical boards dry until your next
scheduled maintenance. Desiccant life span is
dependent on site conditions.

Desiccant Pack Options

Small Desiccant

The Small Desiccant is a disposable cap that ships with
In-Situ Inc. products. The small desiccant is meant to
protect the instrument and cable only during
shipping, and should not be used for
deployments
.

Large Desiccant

The Large Desiccant is a replaceable desiccant pack
designed for low-humidity environments or
deployments where maintenance occurs frequently.
The large desiccant is available with a plastic or a
titanium twist-lock connector.

Information subject to change without notice. In-Situ, In-Situ logo, BaroMerge, BaroTROLL, HERMIT, Pocket-Situ, RDO, RuggedCable,
RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.

©

2012. All rights reserved.

12/2012 | Rev. 002 | 00xxxxx

Advertising
This manual is related to the following products: