In-Situ Quick-Cal Check Solution User Manual

Quick cal solution, Information sheet

Advertising
background image

Information Sheet

Due to continuing product development this information is subject to change without notice. In-Situ, the In-Situ logo, and TROLL are trademarks or regis-
tered trademarks of In-Situ Inc. Copyright © 2002–2006 by In-Situ Inc. All rights reserved.

1 800 446 7488

(toll-free, US and Canada) or 970 498 1500

www.in-situ.com

Multi-Parameter

Water Quality TROLL

®

Catalog No. 33250

0032825 rev. 003 3/06

QUICK CAL
SOLUTION

pH, ORP, & Conductivity values

at various temperatures

pH

mV

µS/cm

Temp.

(± 0.02)

(± 5)

(± 40)

5

o

C

7.10

255

4910

10

o

C

7.04

247

5635

15

o

C

7.03

239

6390

20

o

C

7.02

231

7180

25

o

C

7.00

224

8000

30

o

C

6.98

217

8850

35

o

C

6.97

209

9735

Specific Conductivity Coefficient:

1.95% /

degree C (0-25

o

C)

2.06% /

degree C (15-35

o

C)

2.23% /

degree C (25-50

o

C)

APPLICATION
A calibration standard of an appropriate chemical composition to perform a one-
point calibration of all four MP TROLL Basic sensors (pH, ORP, Conductivity,
Polarographic Dissolved Oxygen) at the same time with a single solution. Four
250 mL containers, each sufficient for at least 2 Quick Cals.

Shelf Life, unopened

6 months

After opening

7-21 days

EXPECTED ACCURACIES
The following accuriacies may be expected when calibrating the MP TROLL
using Quick Cal solution and following instructions in the Win-Situ Quick Cal
Wizard and/or in the MP TROLL Operator's Manual:

GUIDELINES AND PRECAUTIONS

Store the solution cool and dark. Shake well before use. For best results,
calibrate at the temperature expected during field use

Do not store sensors in Quick Cal solution

For best results, use fresh solution for each Quick Calibration

Dispose of used solution by flushing down the drain with plenty of water

Solution is best if used before date on bottle, but may be usable beyond that
date if stored cool, dark, and tightly capped

Quick Cal

Meets USGS

Parameter

accuracy

guidelines*

pH: at pH 7

± 0.09 pH units

over range 0-14 pH

± 0.2 pH units

ORP

± 25 mV

NA

DO (with stirrer)

± 0.25 mg/L

Conductivity:
Low Range (3 – 50,000 µS/cm)

± 3%

High Range (70 – 200,000 µS/cm) ± 3%

* Calibration criteria for continuous water-quality monitors, from USGS Water-Resources
Investigations Report 00-4252, p. 13

Advertising