How to measure, 3 how to measure – KANOMAX 6006 Anemomaster LITE User Manual
Page 11
10
How to Measure
§ 3
How to Measure
- Procedure Flow Chart -
<Change the Time Constant (Air Vel. Measurement Mode)>
By sliding down the switch, you can change the Time Constant
to 1 second (FAST) or to 5 seconds (SLOW).
When the power is turned off, it will return to the initial setting of
1 second (FAST).
When measurement values fluctuate significantly, setting to
“SLOW” will make the reading easy.
<Measurement of Air temperature>
Turn the power on and slide the switch down twice to enter Air
Temperature Measurement Mode.
Do not make a measurement immediately after the mode
change. Especially, when no airflow is present (wind-speed of
0.1m/s or lower), wait for at least 30 seconds before taking a
measurement.
Power is on
Display all items
Air Velocity Measurement
Mode (FAST)
Push and hold the switch for more than 1 second
Slide the switch
down once
Air Temperature
Measurement Mode
Pause
Air Velocity
Measurement Mode
(SLOW)
Slide the switch
up once
Slide the switch
down twice
*When the power is
turned on, air velocity
measurement mode
(FAST) is automatically
selected.
Pause
ON/OFF
Press and hold for
more than 1 second
FAST/SLOW/
Air Temperature
*
Time Constant
determines the time
span of the moving
average. When you
set the Time Constant
to a larger (longer)
value, the readings
will become stable.
When you select a
smaller (shorter)
value, the readings
will become more
responsive and
sensitive to the
change in air velocity.