BIT 4882 030507

5. Adjust control R36 (opposite direction from step 4 above) until the readback value is 0.0

using MEASURE commands (see step 2) as needed. Once a stable value of 0.0 is reached,
continue rotating R36 four full (360°) turns in the same direction. Verify readback current is
0.0 using MEASURE commands (see step 2) as needed.

TABLE 3-2. CALIBRATION MEASUREMENTS AND TOLERANCES - CURRENT

MODEL

SENSE RESISTOR VALUE

(SEE NOTES AND TABLE 3-3)

OUTPUT

CURRENT

ZERO

FULL SCALE

CURRENT

ACCURACY

BOP 100-1M
BOP 200-1M

1 Ohm

0V

±0.000025V

1V (MAX)
-1V (MIN)

±0.000150V

BOP 50-2M

BOP 100-2M

1 Ohm

0V

±0.00005V

2V (MAX)
-2V (MIN)

±0.00030V

BOP 72-3M

1 Ohm

0V

±0.000075V

3V (MAX)
-3V (MIN)
±0.0006V

BOP 50-4

BOP 100-4M

1 Ohm

0V

±0.0001V

4V (MAX)
-4V (MIN)
±0.0008V

BOP 20-5M

0.1 Ohm

0V

±0.0000125V

0.5V (MAX)
-0.5V (MIN)

±0.0000750V

BOP 36-6M
BOP 72-6M

0.1 Ohm

0V

±0.000015V

0.6V (MAX)
-0.6V (MIN)

±0.000090V

BOP 50-8

0.1 Ohm

0V

±0.00002V

0.8V (MAX)
-0.8V (MIN)

±0.00012V

BOP 20-10M

0.1 Ohm

0V

±0.000025V

1.0V (MAX)
-1.0V (MIN)

±0.000150V

BOP 36-12M

0.1 Ohm

0V

±0.00003V

1.2V (MAX)
-1.2V (MIN)

±0.00018V

BOP 20-20M

0.01 Ohm

0V

±0.0000005V

0.2V (MAX)
-0.2V (MIN)

±0.0000030V

REFER TO NOTES IF EXACT
SENSE RESISTOR VALUE IS

KNOWN.

R

SENSE

SEE

NOTE.

SEE

NOTE A.

SEE NOTES C AND

D

USE SPACE PROVIDED AT

RIGHT FOR CALCULATED

VALUES.

NOTES: The following formula is used to calculate proper calibration values if the exact Sense Resistor value is

known.

R

SENSE

=

the measured value of the sense resistor to 6 places (minimum).

I =

Rated current output of BOP (e.g., for BOP 50-4M,

I = 4)

(

R

SENSE

)

(I) (0.00075)

[FOR MODELS WHERE

I = 3

OR

4

USE

(

R

SENSE

)

(I) (0.00012)]

Advertising