B.41 [source:]current:mode command, B.42 [source:]current:mode? query, B.43 [source:]current:protection[:level] command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 1.232 (Unit Rev 0 or 1) User Manual

Page 67: B.44 [source:]current:protection[:level]? query, B.45 [source:]current:protection:state command, B.46 [source:]current:protection:state? query, B.39, Source, B.40, B.41

Advertising
background image

SERIES EL 063010

B-7

B.39 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND

CURR

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP] <NR2>
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] <NR2>
Where <NR2> = current setpoint

Description: Current Setting. Current setpoint is stored in non-volatile memory; value is restored upon power-up.

B.40 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY

CURR?

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns <NR2> = current setpoint

Description: Indicates Current Setpoint.

B.41 [SOURce:]CURRent:MODE COMMAND

CURR:MODE

Syntax:

Short Form: [SOUR:]CURR:MODE

Long Form: [SOURce:]CURRent:MODE

Description: Selects current mode.

B.42 [SOURce:]CURRent:MODE? QUERY

CURR:MODE?

Syntax:

Short Form: [SOUR:]CURR:MODE?

Long Form: [SOURce:]CURRent:MODE?

Returns: <txt> = CURR, POW, VOLT, RES, COND, OFF

Description: Returns the present mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, OFF

B.43 [SOURce:]CURRent:PROTection[:LEVel] COMMAND

CURR:PROT

Syntax:

Short Form: [SOUR:]CURR:PROT[:LEV] <NR2>
Long Form: [SOURce:]CURRent:PROTection[:LEVel] <NR2>
where: <NR2> = current protection setpoint

Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-

ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.

B.44 [SOURce:]CURRent:PROTection[:LEVel]? QUERY

CURR:PROT?

Syntax:

Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns: <NR2> = current protection setpoint

Description: Indicates current protection setpoint.

B.45 [SOURce:]CURRent:PROTection:STATe COMMAND

CURR:PROT:STAT

Syntax:

Short Form: [SOUR:]CURR:PROT:STAT <bool>
Long Form: [SOURce:]CURRent:PROTection:STATe <bool>
where <bool> = 1 (enabled) or 0 (disabled)

Description: Enables (1) or disables (0) the current protection level set using CURR:PROT. This command

saves having to change the protection level to some value outside of the operating range.

B.46 [SOURce:]CURRent:PROTection:STATe? QUERY

CURR:PROT:STAT?

Syntax:

Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: <bool> = 1 (enabled) or 0 (disabled)

Description: Indicates whether the current protection level set using CURR:PROT is enabled (1) or disabled

(0).