Sensor test procedure, Maintenance 4.1. introduction – KROHNE VFM 3100 EN User Manual

Page 29

Advertising
background image

3.5. Sensor Test Procedure

3.5.1. Standard Temperature Range Sensor

Remove electronic module from housing.

Disconnect the yellow and brown sensor leads from back
of electronic module.

Connect sensor lead to an oscilloscope.

With fluid flow in the pipe, observe signal waveform on
oscilloscope. Waveform should be similar to that shown in
Figure 15.

a. If waveform is similar to Figure 15, the sensor is good. If there

is no output from the electronic module, the electronic module
input stage has failed. The entire electronic module should be
replaced.

b. If there is no sensor output signal, the sensor has failed and

should be replaced. See “Sensor Replacement with Integral
Electronic Module” on page 37 for details.

3.5.2. Extended Temperature Range Sensor

Remove the electronic module from the housing. Remove the
preamplifier from the housing. First, pry the ears of the metal
shield away from the sides of the housing. Then lift out the shield
assembly.

Disconnect the brown and yellow sensor leads from the pream-
plifier input terminal strip.

With flow in the pipe, use an oscilloscope to observe the sensor
output. The scope probe impedance must be 10 megohms or
greater. The waveform should be similar to that shown in Figure
15. When the preamplifier is not in the circuit, the minimum
signal required for the sensor is about 2.5 mV.

For liquid flow, the minimum signal of 2.5 mV will require about
25 Hz. Be sure that flow is enough to produce 25 Hz.

For gas or steam flow, the minimum signal of 2.5 mV may require
100 Hz or more, depending on meter size.

If the waveform is similar to Figure 15, the sensor is functioning.
If there is no output, replace the sensor.

NOTE: For all sensors, be sure the signal being read is not the local
power line frequency, i.e., 50 or 60 Hz.

29

Figure 15. Normal VFM 3100 Frequency Waveform

4. Maintenance

4.1. Introduction

The operation of the VFM 3100 consists of three basic functions:
generation and shedding of vortices in the fluid stream, sensing of
vortices, and amplification, conditioning, and processing of the
signal from the vortex sensor. Should a malfunction of the VFM 3100
be suspected, the cause can normally be isolated to one of these
three functions.
Personnel involved in maintenance of VFM 3100 should be trained
and qualified in the use of the equipment required and in the removal
and replacement of the VFM 3100 in the piping. They should also be
qualified for the routine maintenance of the VFM 3100 components.

4.1.1. Vortex Generation and Shedding

The process of vortex generation and shedding can be degraded or
destroyed by disturbances in the upstream flow, the nature of the flo-
wing fluid, or by damage to the vortex shedding element (rare). Such
flow disturbances may be created by gaskets protruding into the flo-
wing stream, by some form of partial blockage in the upstream pi-
ping, by the piping configuration, or by the existence of two-phase
flow. Should the vortex shedding element become heavily caked,
coated, or physically damaged to the extent that its basic shape or
dimensions are changed, the vortex shedding process may be im-
paired. Also, the length of straight, unobstructed run of upstream pi-
ping is important (refer to “Piping Considerations” on page 9).

4.1.2. Vortex Sensing

There are two basic types of sensors employed by the VFM 3100 -
Standard Temperature and Extended Temperature Range. The
Standard Temperature Range sensor may be filled with Fluorolube
Temperature Range sensor is an unfilled sensor for applications to
430 °C (800 °F).
The Standard Temperature Range sensor consists of a piezoelectric
bimorph crystal that is sealed inside a liquid-filled capsule having two
diaphragms on opposing sides. The vortex shedding process
creates an alternating differential pressure across the capsule that is
transmitted via the diaphragms and fill-fluid to the crystal.
The Extended Temperature Range sensor consists of two piezoelec-
tric crystals sealed inside a capsule having two process diaphragms,
on opposing sides, connected internally by a mechanical shuttle.
The vortex shedding process creates an alternating differential
pressure across the capsule that is transmitted via the diaphragm-
shuttle mechanism to the crystals.
The alternating force acting on the crystals causes them to develop
a pulsed voltage with a frequency equal to the vortex shedding
frequency. Damage to sealing diaphragms or other physical damage
could cause the sensors to operate improperly.

4.1.3. Amplification, Conditioning and Processing

The vortex sensor signal is amplified, conditioned, and processed in
the electronic module, which is located in the compartment of the el-
ectrical housing. In addition, the electronic module generates digital,
4 to 20 mA, and scaled pulse output signals. A simplified block dia-
gram of the VFM 3100 is shown in Figure 16.
As shown, the electronic module accepts the raw sensor output
directly from Standard Temperature Range sensors. When used with
an Extended Temperature Range sensor, the raw sensor output
must be buffered by a preamplifier before being passed to the
electronic module. The preamplifier is also used with the standard
temperature range sensor in remote installations. A switch on the
preamplifier is used to match impedance to the sensor. In either ca-
se, the electronic module receives and processes the vortex signal
and provides the various output signals.

Advertising