Installation, 5 wall support for the remote version – KROHNE OPTIWAVE 5200 C_F EN User Manual

Page 34

Advertising
background image

3

INSTALLATION

34

OPTIWAVE 5200 C/F

www.krohne.com

07/2013 - 4001904902 - HB OPTIWAVE 5200 R02 en

Bypass chambers
Installation next to tanks containing one liquid and foam

• The top process connection of the bypass chamber must be above the maximum level of

liquid.

• The bottom process connection of the bypass chamber must be below the lowest measured

level of liquid.

Installation next to tanks containing more than one liquid

• The top process connection of the bypass chamber must be above the maximum level of

liquid.

• The bottom process connection of the bypass chamber must be below the lowest measured

level of liquid.

• Additional process connections are necessary for the liquids to circulate freely along the

length of the bypass chamber.

3.5.5 Wall support for the remote version

1 Use marks on the wall to help you put the wall support in the correct position. For more data,

refer to

Dimensions and weights

on page 116.

2 Use equipement and tools that agree with health and safety regulations and good engineering

practice.

3 Make sure the wall support is correctly attached to the wall.

Figure 3-20: Wall support for the remote version (attached to the remote converter)