KROHNE OPTISOUND 3000 Series EN User Manual
Page 18
Simulation
Simulation
start?
With the reset function, modified values are reset. Three
subfunctions are available:
l
Basic adjustment
- Reset of the values modified with the indicating/
adjustment module to the sensor-specific basic
setting (see chart)
l
Factory setting
- Like basic setting, but also reset of special
parameters modified with PACTware™ to deli-
very status
l
Peak values measured value and temperature
- Reset of the min./max. values of level and
temperature to the current values
Reset values basic setting
Menu
Menu item
Ultrasonic
Basic setting
Units of measurement
deleted
Min. adjustment
upper dead zone depending on instru-
ment
Max. adjustment
end nominal measuring range
Linearization curve
linear
Sensor-TAG
Sensor
Display
Displayed value 1
Distance
Scaling
0% = 0.0, 100% = 100.0
Service
Current output
Output mode: 4-20mA
Failure mode: < 3.6 mA
min. current3.8 mA
Reset
18
Indicating/adjustment module for OPTISOUND sensors
Set up
30603
-EN
-050401