3 instrument repair – KROHNE OPTISWITCH 5200C Transistor EN User Manual

Page 24

Advertising
background image

l

Wrong mode selected

à

Set the correct mode on the mode switch (max.: overfill

protection; min.: dry run protection). Wiring should be

carried out according to the quiescent current principle.

?

Signal lamp flashes red

l

Electronics has detected a failure

à

Exchange instrument or return instrument for repair

?

The signal lamp flashes alternately red and green

l

instrument defective

à

Exchange instrument or return instrument for repair

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our website http://www.

krohne-mar.com/fileadmin/media-lounge/PDF-Download/

Specimen_e.pdf.
By doing this you help us carry out the repair quickly and

without having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and possibly also a safety data

sheet to the instrument