7

TECHNICAL DATA

38

OPTITEMP TT 30 C/R

www.krohne.com

11/2012 - 4000753302 - MA OPTITEMP TT 30 C/R R02 en

Protection category

Protection category

Protection category

Protection category

In-head transmitter

Housing: IP50

Terminals: IP10

Rail-mount transmitter

Housing: IP20

Terminals: IP20

Installation conditions

Mounting

In-head transmitter: "B connection head" or larger according to DIN 43729; with the

help of the rail mounting kit you can also fix this transmitters on a DIN rail

according to DIN 50022 / EN 60715 ( refer to

Rail mounting kit for in-head

transmitters on page 15).
Rail-mount transmitter: rail according to DIN 50022 / EN 60715, 35 mm / 1.38".

For detailed information refer to chapter "Installation".

Weight

In-head transmitter (Ex and Non-Ex version): 50 g / 0.11 lb

Rail-mount transmitter (Ex and Non-Ex version): 70 g / 0.15 lb

Dimensions

For detailed information refer to

Dimensions on page 41.

Materials

Housing

In-head transmitter: PC + ABS, Polyamide for all versions

Rail-mount transmitter: PC + Glass fibre for all versions

Flammability acc. to UL

In-head transmitter: V0 (PC + ABS) and V2 (Polyamide)

Rail-mount transmitter: V0 (PC + glass fibre)

Electrical connections

Power supply

Power supply

Power supply

Power supply

In-head transmitter

Non-Ex version: 6.5...36 VDC (2-wire connection)

Ex version: 8...30 VDC (2-wire connection)

Rail-mount transmitter

Non-Ex version: 7.5...36 VDC (2-wire connection)

Ex version: 8...30 VDC (2-wire connection)

Current consumption

≤ 21,6 mA

Permissible ripple

4 V p-p at 50/60 Hz

Further data

Further data

Further data

Further data

Galvanic isolation

1500 VAC, 1 min

All in- and outputs are electrically isolated from each other and from all other

circuits.

Connection

Single/stranded wires: max. 1.5 mm

2

/ AWG 16

Polarity protection

Standard for all versions

.book Page 38 Friday, November 9, 2012 3:50 PM