TECHNICAL DATA

40

OPTISYS CL 1100

www.krohne.com

01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en

Ozone measurement

Ozone (O

3

) is an instable molecule of three oxygen atoms and a very strong oxidizing agent. At

room temperature it is a gas. Due to its instability it cannot be stored in pressurised cylinders
and has to be prepared on site.

O

3

is an eco-friendly disinfectant. However, its great disinfection strength can only be used to

good advantage in suitable reactors with a sojourn time of at least 3 minutes. The long-term
effect of O

3

is only a few minutes.

O

3

is measured potentiostatic with measuring and counter electrodes of pure gold and an

Ag/AgCl reference. The measurement shows good selectivity towards ozone. A precise potential
is applied to the measuring electrode, leading to an accumulation of negative charges on the
metal surface. O

3

molecules that hit the surface take a defined portion of the charge with them.

The controller measures the potential between measuring and reference electrode and
readjusts the charge on the electrode surface. The current necessary to maintain a constant
charge is a direct measure for the concentration of O

3

. The sensor design with 3 electrodes in a

single rod enables us to use our patented cleaning procedure ASR (patent Dr. A. Kuntze)
providing you with a low-maintenance measuring setup.

Ozon measurement

Figure 7-4: Ozon measurement

1 Reference electrode
2 Voltage measurement
3 Current needed to maintain the constant potential
4 Counter electrode
5 Measuring electrode