The PE (or FE) conductor must be connected to the press-fitted M5 clamp terminal marked
inside the terminal compartment. This conductor must be guided through the rectangular
opening in the metal dividing plate that separates the non-intrinsically safe power supply
terminals from the intrinsically safe signal in-/output terminals.

2.3.1

Connection diagrams MODIS

Section 7 shows the block diagram of the IFM 4080 K/ i…-EEx magnetic-inductive compact
flowmeter. The power supply (terminals 1L½, 0L½) is connected via cable B. The PE terminal
must be connected to the protective earth conductor of the mains supply.

The IFC 090i-EEx electronics unit is provided with intrinsically safe signal in-/output
circuits
due to the installed pair of MODIS modules in accordance with the table below.

Module

Terminal designation

Function / Intrisically safe maximum data

P-SA

I

,

I

Current output (0/4-20 mA), passive
U

i

= 30 V, I

i

= 250 mA, P

i

= 1.0 W

C

i

= 5 nF, L

i

0

FA-ST

B1, B1

or B2, B2

Pulse (frequency) output or status in-/output, all passive
The function can be set by software

U

i

= 30 V, I

i

= 250 mA, P

i

= 1.0 W

C

i

= 5 nF, L

i

0

F-PA

D, D

Fieldbus module, type Profibus system, passive

U

i

= 30 V, I

i

= 300 mA, P

i

= 4.2 W

C

i

= 5 nF, L

i

0

F-FF

D, D

Fieldbus module, type Fieldbus Foundation, passive

U

i

= 30 V, I

i

= 300 mA, P

i

= 4.2 W

C

i

= 5 nF, L

i

0

DC-I

(see note)

I

+, B1+

Intrinsically safe voltage source for the passive module
P-SA or FA-ST, so that active operation is possible.
U

o

= 23.5 V, I

o

= 98 mA, P

o

= 0.6 W

C

o

= 132 nF

1

, L

o

= 4 mH

Table 7: Overview of MODIS modules.

NOTES:

Besides the shown intrinsically safe maximum values for voltages and current -which are
based on certain fault conditions as prescribed by the standard EN 50 020 - the nominal
values for current and voltage must also be respected otherwise a proper funtioning of
the modules is not guaranteed! See table 8 for the nominal values.

The active module DC-I is needed in the 24 Vac/dc power supply version to form an
active current or pulse output in combination with one of the passive modules P-SA or
FA-ST. Due to limited space it is not available for 100...230 Vac supply versions.

Table 9 shows the possible combinations of the installed MODIS modules for the 24 Vac/dc
power supply version of the IFC 090i-EEx and table 10 for the 100-230 Vac version.

1

When modules P-SA (or FA-ST) and DC-I are connected in series, the internal capacitance C

i

of 5 nF must be

subtracted from the C

o

of 132 nF. So the data plate will list a C

o

of 127 nF.