KROHNE UFM 600T EN User Manual

Page 50

Advertising
background image

replacements. If 3 or 4 markers
remain the selection of a diffe-
rent location on the pipe wall is re-
commended; some irregularities at
the interior of the pipe might have
caused poor signal reception.

5

Low signal marker: 1 or 2 markers on.

Cause

Solution

Small deviation of the

Adjust one transducer approx.

transducer distance from ± 5 mm. until the markers dis-
the correct value

appear; do this very carefully so as
not to disturb the sonic coupling
grease layer between the trans-
ducer and the pipe wall.

47

Advertising