OPERATION

34

OPTISENS MAC 080

www.krohne.com

01/2010 - MA MAC 080 R02 en

6.2 Menus of the converter (overview)

Press ↑ + ^ simultaneously for 5 seconds to enter the converter menu.

6.2.1 Menu "Setup"

Menu

Description

Language

Language

Language

Language

English, German or Swedish.

Base unit

Base unit

Base unit

Base unit

Metric or US, selects mm or inch, m

3

or Gallon, °C or °F (the latest

can be overridden in the format menu below).

Date

Date

Date

Date

Show and change current date.

Time

Time

Time

Time

Show and change current time.

Contrast

Contrast

Contrast

Contrast

The contrast is compensated for temperature effects but may have to

be changed due to local lighting conditions or temperature.

Backlight

Backlight

Backlight

Backlight

"Auto" / "On"; Selects if the backlight of the display shall switch of

automatically eight minutes after last key was pressed

(recommended), or be on all the time.

No.Slots

No.Slots

No.Slots

No.Slots

"Auto", "4", "2" or "1"; Selects how many slots shall be shown in the

main menu. "1" shows only slot one, "2" shows slot one and two. "4"

shows all four slots. Auto selects the lowest possible from 1, 2 and 4.

Alarm type

Alarm type

Alarm type

Alarm type

"Level" / "Confirmed"; Select if an alarm shall go away when the level

is OK, or if it shall stay on until confirmed.

Formats

Formats

Formats

Formats

Press ^ to go to the "Formats" sub menu.

Temp

°F or °C.

Date

YY-MM-DD, MM/DD/YY or DD-MM-YY

Time

HH:MM:SS or HH:MM

Exp.Module

Exp.Module

Exp.Module

Exp.Module

Installed module type, press ^ to set the node address if applicable.

Outputs

Outputs

Outputs

Outputs

Press ^ to go to the "Outputs" sub menu.

On fail

"Freeze" / "2 mA"; Selects what shall happen with the analog output

signals if communication to a sensor is lost.

Out sig 1

4...20 mA or 20...4 mA

Out sig 2

4...20 mA or 20...4 mA

Out sig 3

4...20 mA or 20...4 mA (only with extra 4...20 mA module)

Out sig 4

4...20 mA or 20...4 mA (only with extra 4...20 mA module)

Integrate

Integrate

Integrate

Integrate

"Normal" / "Smart"; Smart shortens the integration time when more

than five consecutive samples are on the same side of the average to

get faster response to large changes.

Serial log

Serial log

Serial log

Serial log

"Off", "Slot 1 - Slot 4", "All" or "Slot1&2"; Selects a sensor to be

logged using the serial RS232 interface. The protocol is clear text.

This function is implemented for our service and may change in the

future without notice.

Interv.min

Interv.min

Interv.min

Interv.min

0...999 minutes interval logging a value on the serial channel.

Choosing 0 will give a log value every second.